货号
FY-EH2458S样本类型
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50uL检测时间
1.5H
Contact supplier to negotiate price details
100% Quality Assurance
Fast Shipping
| Concentration | OD1 | OD2 | Average OD | Corrected |
|---|---|---|---|---|
| 500 | 2.472 | 2.502 | 2.487 | 2.45 |
| 250 | 1.616 | 1.627 | 1.622 | 1.585 |
| 125 | 0.95 | 0.959 | 0.955 | 0.918 |
| 62.5 | 0.431 | 0.434 | 0.433 | 0.396 |
| 31.25 | 0.222 | 0.222 | 0.222 | 0.185 |
| 15.63 | 0.124 | 0.124 | 0.124 | 0.087 |
| 7.82 | 0.072 | 0.073 | 0.073 | 0.036 |
| 0 | 0.037 | 0.037 | 0.037 | 0 |
| Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
|---|---|---|---|---|
| 1:2 | Range (%) | 95-103 | 85-94 | 92-101 |
| Average (%) | 96 | 89 | 94 | |
| 1:4 | Range (%) | 86-92 | 87-101 | 87-98 |
| Average (%) | 89 | 88 | 87 | |
| 1:8 | Range (%) | 89-97 | 86-97 | 85-92 |
| Average (%) | 95 | 88 | 92 | |
| 1:16 | Range (%) | 91-103 | 88-95 | 80-96 |
| Average (%) | 93 | 93 | 89 |
| Item | Intra-assay Precision | Intra-assay Precision |
|---|---|---|
| Sample | 3 | 3 |
| Replicate | 9 | 18 |
| CV(%) | 5 | 8 |
| Sample Type | Range(%) | Average Recovery(%) |
|---|---|---|
| Serum (n=8) | 80-92 | 85 |
| EDTA plasma (n=8) | 85-99 | 92 |
| Cell culture media (n=8) | 87-99 | 93 |